Femoroacetabular impingement osteoplasty is any resected amount safe? A laboratory based experiment with sawbones

journal contribution
posted on 2015-09-01, 00:00 authored by B W Loh, C M Stokes, Ben Miller, Richard PageRichard Page
There is an increased risk of fracture following osteoplasty of the femoral neck for cam-type femoroacetabular impingement (FAI). Resection of up to 30% of the anterolateral head–neck junction has previously been considered to be safe, however, iatrogenic fractures have been reported with resections within these limits. We re-evaluated the amount of safe resection at the anterolateral femoral head–neck junction using a biomechanically consistent model.

In total, 28 composite bones were studied in four groups: control, 10% resection, 20% resection and 30% resection. An axial load was applied to the adducted and flexed femur. Peak load, deflection at time of fracture and energy to fracture were assessed using comparison groups.

There was a marked difference in the mean peak load to fracture between the control group and the 10% resection group (p < 0.001). The control group also tolerated significantly more deflection before failure (p < 0.04). The mean peak load (p = 0.172), deflection (p = 0.547), and energy to fracture (p = 0.306) did not differ significantly between the 10%, 20%, and 30% resection groups.

Any resection of the anterolateral quadrant of the femoral head–neck junction for FAI significantly reduces the load-bearing capacity of the proximal femur. After initial resection of cortical bone, there is no further relevant loss of stability regardless of the amount of trabecular bone resected.

Based on our findings we recommend any patients who undergo anterolateral femoral head–neck junction osteoplasty should be advised to modify their post-operative routine until cortical remodelling occurs to minimise the subsequent fracture risk.

History

Journal

Bone & joint journal

Volume

97-B

Issue

9

Pagination

1214 - 1219

Publisher

British Editorial Society of Bone and Joint Surgery

Location

London, Eng.

ISSN

2049-4394

Language

eng

Publication classification

C Journal article; C1.1 Refereed article in a scholarly journal

Copyright notice

2015, British Editorial Society of Bone & Joint Surgery